Skip to content

Hamza et al. 2025 — Physicochemical properties, heavy metals, and micronutrients of Nigerian energy drinks

Hamza and colleagues (Gombe State University and Yobe State University, Nigeria) quantified six heavy metals (cobalt Co, chromium Cr, cadmium Cd, arsenic As, nickel Ni, lead Pb), four essential micronutrients (copper Cu, iron Fe, manganese Mn, zinc Zn), and four physicochemical parameters (pH, turbidity, total dissolved solids TDS, conductivity) in thirty commercially available energy drinks purchased from Nigerian local markets, comprising twenty-three liquid formulations and seven powdered formulations. Liquid samples were first dried by placing a 10 mL aliquot in an oven at 105 °C to remove moisture; the residue was ground, and 1 g was digested with aqua regia (HCl:HNO₃ in a 3:1 ratio) on a Kjeldahl heater for 4–5 hours (no digestion temperature stated by the source), then diluted to 100 mL with deionized water and analysed by atomic absorption spectrophotometry (Bulk 205 AAS). Powdered samples (3 g per sample) were pressed into 25 mm pellets with an automatic hydraulic press, covered with 6 µm polypropylene X-ray foil, and analysed by energy-dispersive X-ray fluorescence (EDXRF). Statistical analyses included Pearson correlation, chi-square against WHO drinking-water comparator values, an independent-samples t-test comparing liquid versus powdered formulations, and two multiple-regression models (Mn ~ Cu + Fe + Zn; Zn ~ Cu + Fe + Mn). The authors report that 43 % of samples exceeded the WHO drinking-water TDS limit of 500 mg L⁻¹, all thirty samples exceeded the WHO conductivity limit of 400 µS cm⁻¹, and all thirty samples were below the WHO pH range of 6.5–8.5 (i.e., acidic); among the heavy metals, lead, cadmium, chromium, and cobalt exceeded the cited WHO drinking-water limits in multiple samples, with the powdered EJ and KR samples showing the highest Pb values (0.2092 and 0.1754 mg L⁻¹ respectively).

Key numbers

All concentrations are reported in the source as mg L⁻¹ on a sample-as-analysed basis. The authors used WHO drinking-water reference values as comparators throughout (Cu 2.0, Fe 0.3 [text]/2.0 [Table 5], Mn 0.4 [text]/0.4 [Table 5], Zn 3.0 [Table 5]/5.0 [text] mg L⁻¹ for micronutrients; Co 0.05, Cr 0.05, Cd 0.003, As 0.01, Ni 0.07, Pb 0.01 mg L⁻¹ for heavy metals). Note the within-paper inconsistency between the chi-square Table 5 expected values and the discussion-text WHO comparator values for Fe and Zn, recorded in Verification notes. The source does not declare an instrument limit of detection; values reported as “ND” (Not Detected) appear across multiple sample/analyte combinations and are treated here as left-censored at the unspecified instrument detection floor.

Physicochemical properties of the thirty energy drinks (Table 1, p. 216)

SampleTypepHTurbidity (NTU)TDS (mg L⁻¹)Conductivity (µS cm⁻¹)
SYLiquid4.531324921036
RBLiquid4.69859741983
PHLiquid4.73957861587
PWLiquid3.06111293416
XCLiquid5.03876881420
HSLiquid2.92110306440
3HLiquid4.52107502718
WBLiquid4.0681386543
BRLiquid3.0168362521
HDLiquid3.971087631092
BHLiquid2.96112412582
ORLiquid4.0253546853
SDLiquid2.9696336347
BSLiquid3.68121582536
MELiquid4.2874460738
VELiquid3.5358524552
FLLiquid3.2598625721
PRLiquid4.9954354472
SKLiquid4.2072462664
IPLiquid3.8781483735
MPLiquid3.0645825592
ARLiquid3.5055471746
CXLiquid3.4238352425
EJPowder4.25746201648
KRPowder5.3682546825
KKPowder4.8488538748
PSPowder5.28648302170
PEPowder5.865810722230
ALPowder4.648420887
ESPowder5.53148431752
WHO (2004)6.5–8.5≤5≤500≤400

Per-parameter summary across the panel: pH ranged 2.92 (HS) – 5.86 (PE); all thirty samples below the WHO 6.5–8.5 lower bound. Turbidity ranged 8 (AL) – 132 (SY) NTU; all thirty samples exceeded the WHO ≤5 NTU limit. TDS ranged 293 (PW) – 1072 (PE) mg L⁻¹, with 43 % of samples exceeding the WHO ≤500 mg L⁻¹ limit (per the source’s discussion text). Conductivity ranged 347 (SD) – 2230 (PE) µS cm⁻¹; all thirty samples exceeded the WHO ≤400 µS cm⁻¹ limit.

Concentration of heavy metals in the thirty energy drinks (Table 2, p. 217–218, mg L⁻¹)

“ND” = Not Detected (left-censored at the unspecified instrument detection floor).

SampleTypeCoCrCdAsNiPb
SYLiquid0.02150.0641ND0.0023ND0.1393
RBLiquidND0.04630.01270.00040.08750.0451
PHLiquid0.0054ND0.0055ND0.0136ND
PWLiquidND0.01750.0131ND0.0652ND
XCLiquidNDNDNDNDND0.0545
HSLiquidND0.09830.01550.00020.0156ND
3HLiquid0.00530.01250.01250.00160.05570.0825
WBLiquidND0.02530.0116ND0.0982ND
BRLiquidND0.04930.0162ND0.0365ND
HDLiquidND0.04730.00840.00370.0625ND
BHLiquid0.00520.25630.0182ND0.0478ND
ORLiquidND0.02630.0198ND0.0984ND
SDLiquid0.00840.06720.0282ND0.0794ND
BSLiquid0.01820.08540.01450.00420.04320.0615
MELiquid0.00290.03230.00740.00280.0608ND
VELiquid0.0165NDND0.00110.0064ND
FLLiquid0.00270.0063ND0.00210.0942ND
PRLiquid0.08260.01130.0106ND0.04510.0451
SKLiquidND0.0046ND0.00120.0075ND
IPLiquid0.01620.02960.0015ND0.0516ND
MPLiquid0.00170.41590.05660.00560.0062ND
ARLiquidND0.02910.00370.00140.0015ND
CXLiquid0.0126NDND0.00360.0183ND
EJPowder0.0835ND0.01280.04510.06240.2092
KRPowder0.05340.3764ND0.03160.00460.1754
KKPowder0.01750.02650.00520.02490.05730.0154
PSPowder0.0263ND0.0074ND0.0573ND
PEPowderND0.0034ND0.00120.06320.1225
ALPowder0.00410.0025ND0.00250.04320.0832
ESPowderND0.25010.01830.00630.0473ND
WHO Limit0.050.050.0030.010.070.01

Per-analyte detected ranges across the panel (Table 2 as authoritative):

  • Co: 0.0017 (MP) – 0.0835 (EJ powder) mg L⁻¹; cobalt exceeded the cited WHO 0.05 mg L⁻¹ comparator in three samples: EJ (0.0835 powder), PR (0.0826 liquid), and KR (0.0534 powder).
  • Cr: 0.0025 (AL powder) – 0.4159 (MP liquid) mg L⁻¹; chromium exceeded the cited WHO 0.05 mg L⁻¹ comparator in multiple samples, with MP (0.4159), KR (0.3764 powder), BH (0.2563), and ES (0.2501 powder) showing the highest values. No speciation was performed; values are total chromium.
  • Cd: 0.0015 (IP) – 0.0566 (MP) mg L⁻¹; cadmium exceeded the cited WHO 0.003 mg L⁻¹ comparator in essentially all samples in which it was detected, with MP (0.0566) showing the highest value.
  • As: 0.0002 (HS) – 0.0451 (EJ powder) mg L⁻¹; total arsenic (no iAs/tAs speciation) exceeded the cited WHO 0.01 mg L⁻¹ comparator in the two highest-concentration powdered samples EJ (0.0451) and KR (0.0316) and KK (0.0249).
  • Ni: 0.0015 (AR) – 0.0984 (OR) mg L⁻¹; nickel exceeded the cited WHO 0.07 mg L⁻¹ comparator in several samples including OR (0.0984), FL (0.0942), WB (0.0982), and RB (0.0875).
  • Pb: 0.0154 (KK powder) – 0.2092 (EJ powder) mg L⁻¹ among detected samples; lead exceeded the cited WHO 0.01 mg L⁻¹ comparator in every detected sample, with the powdered EJ (0.2092) and KR (0.1754) showing the highest concentrations.

Concentration of micronutrients in the thirty energy drinks (Table 3, p. 219–220, mg L⁻¹)

“ND” = Not Detected.

SampleTypeCuFeMnZn
SYLiquid0.07201.9616ND13.8875
RBLiquid0.08450.98120.01950.0527
PHLiquid0.01750.00960.02450.0117
PWLiquid0.06135.70420.04260.0785
XCLiquid0.00270.4183ND0.0492
HSLiquid0.04211.56660.03276.5448
3HLiquid0.08521.87250.02130.0725
WBLiquid0.10361.32520.02570.4542
BRLiquid0.07232.08970.03540.0618
HDLiquid0.08321.06250.03932.7614
BHLiquid0.06592.27910.02740.0713
ORLiquid0.10383.18620.02840.4547
SDLiquid0.08254.51590.02520.2351
BSLiquid0.07361.57560.01630.0492
MELiquid0.73380.06220.54320.0471
VELiquid0.02120.01760.00430.0165
FLLiquid0.08180.0155ND0.0245
PRLiquid0.21610.30430.84420.0109
SKLiquid0.08180.01330.07270.0104
IPLiquid0.79310.52750.26490.0933
MPLiquid0.50130.47250.63330.0726
ARLiquid0.01740.05320.04080.0186
CXLiquid0.00450.01260.01140.0329
EJPowder0.10270.54420.02450.3523
KRPowder0.58410.38610.62180.7421
KKPowder0.31050.06710.41260.0561
PSPowder0.05140.16420.46620.1754
PEPowder0.01810.32880.01570.6437
ALPowder0.04150.79910.00730.1726
ESPowder0.06522.27970.02720.0718

Per-analyte detected ranges across the panel:

  • Cu: 0.0027 (XC) – 0.7931 (IP) mg L⁻¹; all samples below the cited WHO 2.0 mg L⁻¹ drinking-water comparator.
  • Fe: 0.0096 (PH) – 5.7042 (PW) mg L⁻¹; the source’s discussion text cites the WHO Fe comparator as 0.3 mg L⁻¹ and notes that “some values exceed the WHO guideline limit of 2 mg/L,” while Table 5 lists the WHO Fe expected value as 2.0 mg L⁻¹ — a paper-internal inconsistency recorded in Verification notes. Numerous samples exceed the 0.3 mg L⁻¹ figure; PW (5.7042), SD (4.5159), OR (3.1862), BH (2.2791), ES (2.2797 powder), and BR (2.0897) exceed the 2.0 mg L⁻¹ figure.
  • Mn: 0.0043 (VE) – 0.8442 (PR) mg L⁻¹; the source’s text cites a 0.4 mg L⁻¹ comparator for manganese intake-related health risk and notes that “manganese intake above 0.4 mg/L may pose health risks.” PR (0.8442), ME (0.5432), MP (0.6333), KR (0.6218 powder), PS (0.4662 powder), KK (0.4126 powder) exceed this figure.
  • Zn: 0.0104 (SK) – 13.8875 (SY) mg L⁻¹; the source’s text cites a 5 mg L⁻¹ WHO drinking-water comparator while Table 5 lists the expected value as 3.0 mg L⁻¹. SY (13.8875) and HS (6.5448) exceed both figures.

Independent-samples t-test for liquid versus powdered samples (Table 4, p. 220–221, partial)

The source presents only four analytes in the Table 4 t-test panel (Co, Cr, Cu, Fe); the remaining analytes are not tabulated for the liquid-versus-powder comparison.

Elementt-valuep-valueSource interpretation
Cobalt (Co)−1.420.197No significant difference between liquid and powder samples
Chromium (Cr)−0.590.571No significant difference between liquid and powder samples
Copper (Cu)−0.220.832No significant difference between liquid and powder samples
Iron (Fe)1.530.140No significant difference between liquid and powder samples

Chi-square test of observed mean versus WHO expected value (Table 5, p. 225)

ElementObserved mean (mg L⁻¹)WHO expected (mg L⁻¹)
Cu0.14222.0
Fe1.35830.3
Mn0.12130.4
Zn1.00373.0

Combined chi-square value 6.98; p = 0.0725 — not statistically significant at α = 0.05 but suggestive at α = 0.10. The source reports χ² = 6.98 and p = 0.0725 without explicitly stating degrees of freedom (df = 3 is the implied value for a four-element comparison, but the wiki page does not assert this as a source-stated figure).

Pearson correlation matrix among the four micronutrients (Fig. 6, p. 225)

CuFeMnZn
Cu1.00−0.220.67−0.11
Fe1.00−0.350.13
Mn1.00−0.13
Zn1.00

Multiple-regression results for Mn ~ Cu + Fe + Zn (OLS, n=27 observations; Snapshot 1, p. 227)

R² = 0.488; adj-R² = 0.422; F(3, 23) = 7.317; p(F) = 0.00129; Durbin-Watson = 2.273; Jarque-Bera 33.082 (p = 6.55 × 10⁻⁸). Per-predictor coefficients: const 0.0927 (SE 0.061, t = 1.513, p = 0.144); Cu β = 0.6861 (SE 0.172, t = 3.994, p = 0.001, 95 % CI 0.331 – 1.041); Fe β = −0.0338 (SE 0.026, t = −1.281, p = 0.213); Zn β = −0.0088 (SE 0.028, t = −0.315, p = 0.756). The source’s prose summarises Cu as a strong, significant positive predictor of Mn (β = 0.6861, p = 0.001) and Fe and Zn as non-significant.

Multiple-regression results for Zn ~ Cu + Fe + Mn (OLS, n=27 observations; Snapshot 2, p. 228)

R² = 0.018; adj-R² = −0.110; F(3, 23) = 0.1392; p(F) = 0.936; Durbin-Watson = 2.285; Jarque-Bera 293.079 (p < 2.28 × 10⁻⁶⁴). Per-predictor coefficients: const 0.6171 (SE 0.460, t = 1.340, p = 0.193); Cu β = −0.3096 (SE 1.662, t = −0.186, p = 0.854); Fe β = 0.0056 (SE 0.203, t = 0.028, p = 0.978); Mn β = −0.4878 (SE 1.548, t = −0.315, p = 0.756). The model does not fit the Zn data.

Methods (brief)

Thirty energy drinks (twenty-three liquid and seven powdered) were purchased from local markets in Nigeria and refrigerated at 4 °C until analysis. Liquid samples were prepared for AAS by transferring a 10 mL aliquot into an oven at 105 °C to remove moisture, grinding the dried residue with a pestle and mortar, and digesting precisely 1 g of ground sample in 20 mL of aqua regia (HCl : HNO₃ 3 : 1) inside a fume hood on a Kjeldahl heater for 4–5 hours with continuous addition of acid until a pale yellow solution indicated complete decomposition. After cooling, the digest was diluted with deionized water, filtered, and made up to 100 mL with deionized water. Metal concentrations were quantified using a Bulk 205 Atomic Absorption Spectrophotometer. The source does not specify whether flame-AAS or graphite-furnace-AAS was used, does not report per-analyte wavelengths, does not report instrument limits of detection or quantification, does not report certified-reference-material recoveries, and does not specify replicate structure per sample. Powdered samples (3 g each) were placed into 25 mm diameter pellet moulds, covered with a 6 µm polypropylene transparent X-ray foil, and compressed into pellets using an automatic hydraulic press; the pellets were loaded into the X-ray excitation chamber of an EDXRF spectrometer using an automated sample changer system, and a time-based irradiation program controlled by dedicated software was used to analyse both samples and standard reference materials (the reference materials are not further specified).

Physicochemical parameters were measured with: a JENWAY 3505 digital pH meter calibrated with pH 4.0/7.0/10.0 buffers; a HACH Sension 5 digital conductivity meter calibrated with a KCl standard; a HACH DR/890 digital turbidity colorimeter calibrated against turbidity standards; and a HACH Sension 5 digital TDS/conductivity meter calibrated with a standard solution.

Statistical analyses comprised Pearson correlation (output shown as a heat-mapped 4 × 4 matrix for the micronutrient analytes), chi-square comparing observed means against WHO drinking-water comparator values (χ² = 6.98, p = 0.0725; the source does not state degrees of freedom), an independent-samples t-test comparing liquid versus powder formulations on Co/Cr/Cu/Fe (none significant at α = 0.05), and two ordinary-least-squares multiple-regression models fitted on n=27 observations (Mn ~ Cu + Fe + Zn; Zn ~ Cu + Fe + Mn). No metal speciation was performed: arsenic is recorded as total arsenic (tAs) per CLAUDE.md Part 14, and chromium is recorded as total chromium (Cr) — no Cr-VI distinction is offered by the source.

Evidence Fitness

This source contributes direct primary occurrence values for ten metals (Pb, Cd, tAs, Cr, Ni, Co, Cu, Fe, Mn, Zn) plus four physicochemical parameters across thirty packaged energy drinks (twenty-three liquid and seven powdered) sold at Nigerian local markets. The principal limitations bearing on pooling eligibility and synthesis weight are:

(i) Same-dataset overlap with the companion Babayo et al. 2026 Nigerian energy-drinks paper (babayo2026-heavy-metals-energy-drinks-nigeria). The thirty sample codes (SY, RB, PH, PW, XC, HS, 3H, WB, BR, HD, BH, OR, SD, BS, ME, VE, FL, PR, SK, IP, MP, AR, CX, EJ, KR, KK, PS, PE, AL, ES) and the Cd/Cr/Co/As/Ni/Pb values in Table 2 of this paper match the corresponding values in the Babayo et al. 2026 paper, where H.A. Hamza is also a co-author. The two papers report the same underlying analytical dataset but differ in downstream analysis: the present paper adds physicochemical parameters, micronutrients (Cu, Fe, Mn, Zn), and multiple-regression analyses; the Babayo paper adds Contamination Factor, Metal Pollution Index, and Estimated Daily Intake calculations. Synthesis must avoid double-counting these as two independent observations of the Nigerian energy-drinks population; the routing layer treats them as a single dataset for pooling. The near_duplicates: frontmatter records the relationship.

(ii) No declared analytical limit of detection or quantification. The “ND” attribution is repeated across the table without an explicit numerical floor. The lack of declared LOD/LOQ prevents pool-eligibility decisions to be made under standard censored-data treatment.

(iii) No certified reference material or recovery data reported for the AAS pathway. The XRF pathway mentions “standard reference materials” loaded into the X-ray excitation chamber but does not name the certified reference material vendor or composition.

(iv) Two distinct analytical methods used on a single dataset (AAS for liquids; EDXRF for powders). Different matrix-effects, sensitivity, and detection-limit characteristics apply to the two methods; pooling liquid and powder values within the same analyte column risks confounding analytical-method effects with sample-type effects. The source’s own t-test (Table 4) reports no significant liquid-versus-powder difference for Co/Cr/Cu/Fe but does not extend the test to Cd/As/Ni/Pb/Mn/Zn.

(v) No metal speciation. Arsenic is reported as total elemental arsenic (recorded as tAs); chromium is reported as total chromium without Cr(VI) speciation; mercury is not in the analyte panel. Among the ten HMTc/HMI analytes (Pb, tAs, Cd, MeHg, tHg, iAs, Ni, Al, Cr-VI, Sn), this source covers Pb, tAs (no iAs), Cd, Ni, and Cr (no Cr-VI) — five of ten in part, with the speciation caveats noted.

(vi) Paper-internal inconsistency in the cited WHO comparator values for Fe and Zn. The discussion text (p. 219) cites WHO Fe at 0.3 mg L⁻¹ (and also notes “some values exceed the WHO guideline limit of 2 mg/L”) and WHO Zn at 5 mg L⁻¹, while Table 5 lists Fe expected at 2.0 mg L⁻¹ and Zn expected at 3.0 mg L⁻¹. The exceedance counts in this wiki page describe both comparator versions; downstream synthesis should resolve which WHO guideline edition the authors intended.

(vii) Tables 2 and 3 list units as “mg/L” for both liquid and powdered samples. For powdered samples this implicitly assumes a reconstitution basis that the source does not state. The values are recorded here as the source presents them; downstream pooling against an as-consumed liquid basis should treat powdered values cautiously until the reconstitution dilution is verified.

(viii) Multiple-regression sample size mismatch (n=27 reported in the OLS output; n=30 stated in the methods). Snapshots 1 and 2 both report “No. Observations: 27” while the dataset contains thirty samples; the source does not state how the three missing observations were dropped (likely listwise deletion against the Mn=ND values in SY/XC/FL and the Zn=ND value in HS — but the source does not say). This caveat affects only the regression results, not the raw concentration tables.

Evidence tier set to C. The source is primary research, peer-reviewed (Communication in Physical Sciences, Vol 12, Issue 3, March 2025), but with the methodological caveats above (no declared replicate structure, no declared LOD/LOQ, no CRM recovery data for AAS, mixed-method dataset, no metal speciation, internal-numerics inconsistencies, and the powder-basis ambiguity). The raw concentration values in Tables 2 and 3 are pool-considerable for Pb, Cd, tAs, Cr, Ni, Co, Cu, Fe, Mn, and Zn at C-tier weight with explicit notation of caveats (i)–(viii), and should be co-pooled with the Babayo companion paper as a single dataset rather than as two independent observations.

Implications

  • Certification: contributes direct primary occurrence values for the sports-energy-drinks HMTc category (Category 5) for the Nigerian retail market. The Pb column reports eleven detected samples ranging 0.0154–0.2092 mg L⁻¹ against the source’s cited WHO 0.01 mg L⁻¹ comparator; the Cd column reports twenty-one detected samples ranging 0.0015–0.0566 mg L⁻¹ against the cited 0.003 mg L⁻¹ comparator; the Cr column reports twenty-four detected samples ranging 0.0025–0.4159 mg L⁻¹ against the cited 0.05 mg L⁻¹ comparator; the As column reports nineteen detected total-arsenic samples ranging 0.0002–0.0451 mg L⁻¹ against the cited 0.01 mg L⁻¹ comparator. These values are co-pooled with the companion Babayo et al. 2026 paper (babayo2026-heavy-metals-energy-drinks-nigeria) and complement the prior Bunu et al. 2023 Kogi-State (bunu2023-heavy-metals-energy-drinks-kogi) and Bayelsa-State (bunu2023-heavy-metals-energy-drinks-bayelsa) Nigerian data, plus the Polish-market Czarnek et al. 2024 (czarnek2024-heavy-metals-energy-drinks) and Jordanian-market Al-Sayyed et al. 2024 (alsayyed2024-heavy-metals-energy-drinks-jordan) comparators. The two Hamza/Babayo Gombe-State-University dataset publications together (n=30, 23 liquid + 7 powdered) form a single observation block for the Nigerian energy-drinks occurrence picture.
  • Courses: useful as a teaching reference for (1) the importance of declared LOD/LOQ and CRM recoveries for trace-metal analysis of finished beverages; (2) the importance of separating analytical-method effects (AAS vs EDXRF) from sample-type effects (liquid vs powder) when pooling mixed-method datasets; (3) the importance of speciation (Cr vs Cr-VI; tAs vs iAs) for downstream toxicology interpretation; (4) the recurring pattern in low-resource analytical reporting where the WHO drinking-water comparator is applied to packaged beverages without addressing the as-consumed-versus-drinking-water basis; (5) the importance of declaring how dropped observations are handled in regression analyses (the n=27 vs n=30 mismatch).
  • App: contributes Nigerian-market occurrence values for the packaged energy-drink product class on a single-dataset, sample-as-analysed basis. Per-sample brand identities are not disclosed by the source.
  • Discovery: the source’s reference list overlaps substantially with the Babayo companion paper and includes earlier Nigerian-beverage heavy-metals surveys (Iweala et al. 2014, Izah et al. 2015, Maduabuchi et al. 2006, Onianwa et al. 1999, Salako et al. 2016, Obuzor & Ajaezi 2010); the highlighted “Jingjing L., Qianlu Q. & Yanshan C. (2020). Arsenic speciation in energy drinks and human urine by HPLC-IC-MS. Journal of Agriculture and Food Chemistry, 68, 2” and “Mehmet O., John D. R. & Jennifer L. M. (2019). Arsenic in energy drinks: A risk assessment. Journal of Food Science, 84, 5” references (highlighted in yellow in the source PDF) are potential As-speciation discovery leads.

Provenance notes

Open-access publication in Communication in Physical Sciences (CPS), Vol 12, Issue 3, March 2025, pp. 213–230. DOI: 10.4314/cps.v12i3.19. Received 18 January 2025; Accepted 02 March 2025; Published 14 March 2025. Corresponding author: Hamza Abubakar Hamza, Department of Pure and Applied Physics, Gombe State University, P.M.B. 127, Gombe, Nigeria (hamzadurbi@gsu.edu.ng; ORCID 0009-0005-9958-5508). Co-authors: Abubakar Danjuma Bajoga (Gombe State University; a.bajoga@gsu.edu.ng), Yusuf Mohammed Auwal (Yobe State University, Damaturu; auwalgamajan@ysu.edu.ng), Hankouraou Seydou (Gombe State University; seydou5k@yahoo.com). Funding: no external source of funding declared. Conflict of interest: none declared. Ethical considerations: not applicable (no human or animal subjects). Data availability: stated as “available on request.” Accessed via the Manual Fetch Discovery autopilot.

Wiki pages this source may touch

Verification notes

Speciation handling per CLAUDE.md Part 14. This source measures total elemental Pb, Cd, As, Cr, Ni, Co, Cu, Fe, Mn, and Zn by AAS (liquid) and EDXRF (powder), without speciation. Arsenic is recorded as tAs (total arsenic) in the metals: frontmatter field, per CLAUDE.md Part 14’s “iAs/tAs and tHg/MeHg as non-negotiable distinctions” rule. Chromium is recorded as Cr (total chromium); no Cr-VI distinction is offered by the source, so Cr-VI is NOT recorded. Mercury, aluminium, tin, antimony, and uranium were not in the source’s analyte panel and are not recorded. The micronutrients Cu, Fe, Mn, and Zn are recorded in the metals: field because their values are part of the source’s analytical dataset; they are not HMTc certification analytes but are HMI taxonomy-recognised metal slugs.

Products frontmatter. The products: frontmatter lists only sports-energy-drinks because the source’s thirty-sample panel is described uniformly as “energy drinks” in title, abstract, methods, and results, without mixed-class language. This is in contrast to the Bunu et al. 2023 Kogi-State and Bayelsa-State companion papers, which mix carbonated soft drinks and energy drinks under the “soft drinks” body framing but title themselves “Energy Drinks.”

Matrices frontmatter. The matrices: field uses both energy-drinks (for the twenty-three liquid samples) and energy-drink-powder (for the seven powdered samples), matching the established vocabulary used in the companion Babayo et al. 2026 paper.

Jurisdictions frontmatter. NG (Nigeria). The source does not narrow the sampling to a specific Nigerian state — the methods section simply says “purchased from local markets” without further geographic detail.

Near-duplicate handling per the same-dataset overlap with Babayo et al. 2026. The thirty sample codes and heavy-metal concentration values in Table 2 of this paper match the corresponding values in the Babayo et al. 2026 Nigerian energy-drinks paper (babayo2026-heavy-metals-energy-drinks-nigeria), where H.A. Hamza is also a co-author. The two papers share the underlying analytical dataset but differ in downstream analysis: the present paper adds physicochemical parameters (pH, turbidity, TDS, conductivity), the micronutrients (Cu, Fe, Mn, Zn), and multiple-regression analyses; the Babayo paper adds Contamination Factor, Metal Pollution Index, and Estimated Daily Intake calculations. Recorded in the near_duplicates: frontmatter so that the synthesis pass treats them as a single observation block for pooling.

Brand firewall per CLAUDE.md Part 12 (strict reading, locked 2026-05-17). The source labels samples only by two-letter codes (SY, RB, PH, PW, XC, HS, 3H, WB, BR, HD, BH, OR, SD, BS, ME, VE, FL, PR, SK, IP, MP, AR, CX, EJ, KR, KK, PS, PE, AL, ES) and does not disclose per-code brand mapping anywhere in the text. The two-letter codes are not brand identifiers and are preserved on this wiki page as the source’s neutral sample-labelling convention. The methods section’s scientific-method vendor identities (Bulk 205 Atomic Absorption Spectrophotometer; JENWAY 3505 pH meter; HACH Sension 5 conductivity/TDS meter; HACH DR/890 turbidity colorimeter) are retained under the Part 12 scientific-method-vendor exception locked 2026-05-17.

Wiki/HMTc firewall per CLAUDE.md Part 2. No HMTc threshold proposals, no consumer-audience risk advisories, and no synthesis claims of the form “this confirms the literature consensus that…” appear in this wiki page body. The observation that several samples exceed the cited WHO drinking-water comparator for Pb, Cd, Cr, As, and Ni is reported as the source itself reports it — as a comparison against the regulatory comparator values the source itself cited — and is not framed as an HMTc threshold recommendation or as a consumer-safety claim. The discussion sections of the source contain consumer-audience risk language (“manufacturers should adjust formulations to reduce acidity”; “long-term cadmium exposure can cause kidney damage and skeletal disorders”); these are not reproduced on this wiki page.

Paper-internal inconsistencies recorded faithfully and flagged.

  1. The discussion text (p. 219) cites WHO Fe at 0.3 mg L⁻¹ and Zn at 5 mg L⁻¹, while Table 5 (p. 225) lists the WHO expected value for Fe as 2.0 mg L⁻¹ and for Zn as 3.0 mg L⁻¹. The 0.3 mg L⁻¹ Fe figure is consistent with the WHO drinking-water aesthetic threshold; the 2.0 mg L⁻¹ Fe figure is consistent with an older WHO health-based guideline. The Zn 5.0 mg L⁻¹ figure is consistent with the WHO drinking-water aesthetic threshold; the 3.0 mg L⁻¹ Zn figure does not match any standard WHO drinking-water comparator known to this wiki. Recorded faithfully; downstream synthesis should resolve which guideline edition was intended.
  2. The discussion text (p. 217) reports “43 % of samples exceeding the WHO-recommended limit of 500 mg/L” for TDS; spot-checking against Table 1 yields 13 of 30 samples exceeding 500 mg/L (= 43.3 %), consistent with the source’s text.
  3. The two multiple-regression models report “No. Observations: 27” in the OLS output blocks (Snapshots 1 and 2, pp. 227–228), while the dataset is stated as n = 30 throughout the methods and results. The source does not explain how the three missing observations were dropped. This caveat is restricted to the regression results and does not affect the raw concentration tables.
  4. Figures 1 and 2 in the source carry the same boxplot graphic for copper (Cu) concentration but Figure 2 is captioned as a “Bar Chart of Average Iron (Fe) Concentration in Energy Drinks” — a likely figure-caption-to-graphic mismatch in the published version. Recorded as an observation; does not affect the numerical tables.

Sample size used for tabulated heavy-metal values. Table 2 reports values for all thirty samples (23 liquid + 7 powder). Table 4’s t-test panel covers only four analytes (Co, Cr, Cu, Fe) of the ten analytes measured. Table 5’s chi-square panel covers only the four micronutrients (Cu, Fe, Mn, Zn) and not the six heavy metals; the source does not extend the chi-square test to Pb, Cd, As, Cr, Co, or Ni.

Cite-key choice. hamza2025-heavy-metals-energy-drinks-nigeria rather than including “physicochemical” in the cite-key, for consistency with the existing energy-drinks corpus (alsayyed2024, babayo2026, bunu2023-bayelsa, bunu2023-kogi, czarnek2024). The raw_handle field preserves the auto-generated MFD_hamza2025-physicochemical-heavy-metals-energy-drin handle from the Manual Fetch Discovery autopilot.

Audit application (2026-06-06). Fresh-context Agent subagent audit (verdict REVISE; four ❌ findings on Check 1 numerical fidelity counts in ## Implications plus two ⚠️ concerns on a Co exceedance roll-up and a chi-square df attribution). Independent re-verification against Table 2 (PDF pp. 217–218) and PDF p. 215 confirmed all six findings were correct:

  • Pb detected-sample count corrected from “twelve” to “eleven” (re-count: SY, RB, XC, 3H, BS, PR, EJ, KR, KK, PE, AL).
  • Cd detected-sample count corrected from “twenty-two” to “twenty-one” (re-count: RB, PH, PW, HS, 3H, WB, BR, HD, BH, OR, SD, BS, ME, PR, IP, MP, AR, EJ, KK, PS, ES).
  • Cr detected-sample count corrected from “twenty-three” to “twenty-four” (re-count: 30 samples minus the 6 ND samples PH/XC/VE/CX/EJ/PS = 24).
  • As detected-sample count corrected from “eighteen” to “nineteen” (re-count: SY, RB, HS, 3H, HD, BS, ME, VE, FL, SK, MP, AR, CX, EJ, KR, KK, PE, AL, ES).
  • Co exceedance roll-up corrected to add PR liquid (0.0826 mg L⁻¹ > WHO 0.05) alongside EJ (0.0835) and KR (0.0534).
  • Chi-square df attribution softened from asserted “df = 3” to inferred-not-stated, both in the Table 5 narrative and the Methods (brief) section.
  • The abstract paragraph’s conflation of the 105 °C oven-drying step with the Kjeldahl-heater digestion step was tightened to match PDF p. 215, which states the 105 °C is the oven-drying step and the Kjeldahl heater digestion has no stated temperature. No false-positive findings to record. No frontmatter changes required. Routing unaffected (sports-energy-drinks direct_evidence row unchanged).

Page history

The five most recent substantive edits to this page. The full version history lives in git; when DOI minting comes online (see schema docs), each entry below will also link to a version-pinned DataCite DOI.

CommitDateDescription
e705b0a2026-06-06ingest: hamza2026-heavy-metals-energy-drinks-health-risk fresh from Manual Fetch Discovery
e7cfc9f2026-06-06audit: hamza2025-heavy-metals-energy-drinks-nigeria revised
d3ac7342026-06-06ingest: hamza2025-heavy-metals-energy-drinks-nigeria fresh from Manual Fetch Discovery